Installation
Secure the cable with a cable clip
i
EN
US
Plug the mains plug into the power outlet
i
Replace the cover
i
The cover on the rear must be cor-
rectly fitted when the device is oper-
ated within the patient environment
7.7 Connecting the unit to the
network
Purpose of the network connection
The network connection is used to exchange in-
formation or control signals between the unit
and a software installed on a computer, in order
to, e g:
– Display parameters
– Select operating modes
– Indicate messages and error situations
– Change device settings
– Activate test functions
– Transmit data for archiving
– Provide documents concerning the devices
The unit can be connected to the network with
a network cable or via WLAN
For information on connection via WLAN,
see "81 Installing and configuring the
unit"
Combining devices safely
Safety and essential performance features are
i
independent of the network The device is de-
signed appropriately for operation indepen-
dent of a network However, some of the
functions are not available in this case
Faulty manual configuration can lead to signifi-
i
cant network problems The expert knowl-
edge of a network administrator is required for
configuration
18
OBJ_DOKU-0000002448-001-mehrere.indd 18
1
2
The data connection utilizes part of the band-
i
width of the network Interactions with other
medical devices cannot be completely exclud-
ed Apply the IEC 80001-1 standard for risk
assessment
The device is not suitable to be connected di-
i
rectly to the public internet
Danger can arise when connecting appliances
to each other or to parts of systems (eg
through leakage currents)
Only connect appliances together when there
i
can be no danger to the operator or to the
patient
Only connect appliances when there can be
i
no environmental impairment through such in-
terconnection
If it is not clear from the appliance data sheets
i
that such connections can be safely made or
if you are in any doubt, always get a suitably
qualified person (eg the manufacturer) to ver-
ify that the setup is safe
A template for the system manufacturer's
declaration in accordance with Article 12
of Directive 93/42/EEC can be found in
our download section at wwwduerrden-
talcom (document no 9000-461-264)
Only connect peripheral units (eg computer,
i
monitor, printer) that conform at least to the
requirements set out in IEC 60950-1
(EN 60950-1)
Connecting the unit via the network cable
Remove the cover from the rear of the device
i
Plug the enclosed network cable into the net-
i
work socket of the device
Replace the cover
i
The cover on the rear must be cor-
rectly fitted when the device is oper-
ated within the patient environment
2142100008L41 1708V002
24.08.2017 14:41:45