Electrical Connection; Connecting The Unit To The Network - Air Techniques ScanX Swift View Instrucciones Para Montaje Y Uso

Ocultar thumbs Ver también para ScanX Swift View:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Check whether the memory card has been
inserted correctly in the unit. If the memory is
inserted incorrectly in the device, take it out
again and re-insert it properly.
7.5

Electrical connection

Electrical safety when making connections
Connect the device to a correctly installed
power outlet only.
Do not operate any other systems using the
same multiple socket.
Make sure that none of the electrical cables
leading to the unit are under any mechanical
tension.
Before initial start-up verify that the mains sup-
ply voltage and the voltage stated on the type
plate match (see also "4. Technical data").
Connecting the unit to the mains supply
The unit has no main power switch. For
this reason, it is important to set up the
unit properly such that the plug can be
easily accessed and unplugged if
required.
Requirements:
ü Properly installed power outlet close to the unit
(observe the max. length of the power cord)
ü Easily accessible power outlet
ü Mains voltage matches the information shown
on the type plate of the power supply unit
2142100008L41 2202V007
Attach the matching country-specific mains
cable to the power supply unit.
Remove the cover from the rear of the device.
Plug the connecting plug of the power supply
unit into the connection socket of the unit.
Secure the cable with a cable clip.
2
Plug the mains plug into the power outlet.
Replace the cover.
The cover on the rear must be correctly
fitted when the device is operated
within the patient environment.
7.6
Connecting the unit to the
network
Purpose of the network connection
Installation
1
EN-
US
19
Tabla de contenido
loading

Tabla de contenido