Erection and mounting
The calibration and adjustment facility must be
installed on a level surface with sufficient load-
bearing capacity. It can be secured using the
mounting holes in the handles at the sides. Perfect
alignment and levelling of the calibration and
adjustment facility are important in this case.
To install the calibration and adjustment facility,
place it onto the work surface on the workbench.
Bolt the calibration and adjustment facility to the
work surface using the mounting holes (1).
Check the alignment of the calibration and
adjustment facility.
Make sure that the calibration and adjustment
facility is perfectly aligned and securely mounted.
Installing extension 7791-1
The calibration and adjustment facility is designed
for a maximum torque of 400 N m.
Extension 7791-1 can be used for calibration
objects which require a higher torque. This is
available as an accessory. When extension 7791-1
is installed, a maximum torque of 1100 N m can be
applied.
The extension's scope of delivery contains:
• Extension 7791-1
• Two height compensation plates
• Two bolts for securing to the calibration and
adjustment facility
• Two connectors with tapped holes
A ball-ended hexagon key wrench size 5 is required
to mount extension 7791-1.
Proceed as follows to mount extension 7791-1:
Remove the two cheese head screws with
hexagon socket, to which the handle on the right-
hand side of the calibration and adjustment facility
is secured.
Erection and mounting