Remove the old ex-sensor from the housing and replace with a new ex-
sensor. Shorten the sensor cables on the new sensor to 55 mm, strip
approx. 6 mm and apply the isolated ferrules provided (Zoller+Fröhlich,
type: V3AE0003 or equivalent). This ensures that the degree of protection
IP 30 is preserved if the housing is opened for maintenance purposes.
Screw the new sensor into the hexagon nut (via the housing opening) and
fasten with thread-locking fluid (e.g. Loctite no. 221). To preserve the IP
degree of protection, ensure the flawless fit of the sensor sealing ring.
Connect the ex-sensor's cables (brown, yellow, black) to the corresponding
br/br, ge/yw and sw/bk terminal. Use a screwdriver (width of 3 mm) or
special tool to open the spring terminals (see "Handling the spring terminal
clamps in the transmitter", page 36).
Refit the upper part of the casing, ensuring that the atmosphere is free of
dust and fasten the cover screws.
Turn the transmitter on again and/or re-insert the controller's channel card
into the rack.
Observe the warm-up time of the new ex-sensor:
DrägerSensor PR M DQ, at least 30 minutes
DrägerSensor LC M, at least 120 minutes
After replacing the ex-sensor, the transmitter must be calibrated (see
"Calibrating the transmitter", page 37).