Adapter type number
Battery
Power port
Operating conditions
Storage and
transportation conditions
Operating pressure
Net weight
External dimensions
Device classification
Materials
Service life
Mode of operation
Ingress protection
This breast pump contains a battery that is non-replaceable.
Electromagnetic compatibility - compliance information
This device has been tested and found to comply with IEC 60601-1-2:2014 Ed 4.0 for electromagnetic
compatibility (EMC) as Class B according to CISPR 11:2009. Tests levels are listed in the accompanying
tables. Only use this Philips Avent single/double electric breast pump Advanced in combination with
the adapter supplied.
Single electric breast pump
S009AHz050yyyy The letters "yyyy" represent the output current from
0100 (1000 mA) to 0180 (1800 mA), increments in steps of 100 mA. The
letter "z" denotes type of plug, it can be: U, V, B, S, C, A, D, K, T, I, E, F.
Rechargeable battery: 3.6 V 2600 mAh (non-replaceable)
Micro-USB
Temperature: 5 °C to 40 °C (41 °F to 104 °F)
Relative humidity: 15 to 90 % (non-condensing)
Temperature: -20 °C to 45 °C (-4 °F to 113 °F)
Relative humidity: up to 90 % (non-condensing)
700-1060 hPa (<3000 m/9842 ft. altitude)
230 g
Motor unit: 145 mm x 95 mm x 45 mm (LxWxH)
Applicable classifications:
Degree of protection against electric shock;
- Class II ME equipment
- Type BF applied part
The breast pump and the adapter are specified as medical electrical
system according to IEC 60601-1.
ABS, silicone (motor unit)
Polyester, nylon, polypropylene, rubber (breast pump belt)
Polypropylene, silicone (other parts)
500 hours
Continuous operation
IP22 (motor unit), IP20 (adapter)
Double electric breast pump
310 g
Non-continuous operation: 55 min
on, 180 min off at 22 °C (72 °F) (duty
cycle 23 %) 20 min on, 90 min off at
40 °C (104 °F) (duty cycle 18 %)
English