6.2 Water Line Requirements
• Incoming water supply line pressure to the ultrasonic scaler
must be 25 psi (172kPa) minimum to 60 psi (414 kPa)
maximum. If your dental water system's supply line pressure
is above 60 psi, install a water pressure regulator on the water
supply line to your Ultrasonic Scaler.
• Incoming water temperature to the Cavitron System should not
exceed 25˚C (77˚F). If needed device should be installed to
maintain a temperature within this specification, or a Cavitron
DualSelect Dispensing System attached to allow this system to
be operated as a closed water system.
• A manual shut-off valve on the dental water system supply line
should be used so that the water can be completely shut-off
when the office is unoccupied.
• A filter in the dental water system supply line is recommended
so that the particles in the water supply will be trapped before
reaching the ultrasonic scaler.
• After the above installations are complete on the dental
water supply system, the dental office water line should be
thoroughly flushed prior to connection to the ultrasonic system.
• After flushing system verify there are no leaks.