This product contains metallic
components. Do NOT use in a Magnetic
Resonance (MR) environment.
Precautions
Do not use if the packaging or contents
are damaged. Damage may include
cracks, scratches, dents or any signs
that the monitor, modules, cables,
cable housing or system accessories
may have been compromised during
shipping or unpacking.
HemoSphere Advanced
Monitor Setup
Refer to the HemoSphere advanced
monitor operator's manual included on
the enclosed USB memory stick or
located on the Edwards website
(eifu.edwards.com) for comprehensive
monitoring procedures, cautions and
warnings.
1. Securely position the HemoSphere
advanced monitor or mount on a
compatible stand.
2. Open the battery door on the lower
right panel of the HemoSphere
advanced monitor 
and insert the
battery.
3. Insert desired hemodynamic
technology module 
and/or
cable  . Refer to module or cable
instructions for proper insertion/
connection procedures.
4. Connect the power cord to the rear
panel of the monitor 
into a hospital grade outlet.
Note: After inserting the battery for
the first time, the battery must be
fully charged and conditioned. Refer
to operator's manual for additional
information.
5. Press the power button on the front
of the monitor  .
Note: Upon initial startup the
display language can be selected
which affects the time/date format
and units of measurement.
6. Follow instructions contained in the
HemoSphere advanced monitor
operator's manual for desired
hemodynamic technology module
or cable monitoring.
MRI Information
MR Unsafe
Do not use the HemoSphere advanced
monitor in an MR environment. The
monitor is MR unsafe since the device
contains metallic components, which
can experience RF-induced heating in
the MRI environment.
Operating conditions
Temperature: 10 to 32.5 °C
Humidity range: 20 to 90%
non-condensing
Altitude (Atmospheric Pressure):
0 ft/0 m (1013 hPa) to
3048 m/10,000 ft (697 hPa)
IP Rating: IPX1
Maintenance
The HemoSphere advanced monitor is a
multi-use device and can be surface
cleaned using a cloth dampened with
cleaning agents containing 70%
and plug
isopropyl alcohol, 2% glutaraldehyde,
10% bleach solution, or quaternary
ammonium solution.
Do not use any other cleaning agents,
spray, or pour cleaning solution directly
on the monitor.
Storage
Temperature: -18 to 45 °C
Relative humidity: 90%
non-condensing at 45 °C
Altitude: 0 to 20,000 feet (6096 m)
Note: For long term storage
(>3 months), a maximum storage
temperature of 35 °C is recommended.
Battery capacity starts to degrade with
extended exposure above 35 °C.
Technical Assistance
For technical assistance, please call
Edwards Technical Support at the
following telephone numbers:
Inside the U.S. and Canada
(24 hours): . . . . . . . . . . . . 800.822.9837
Outside the U.S. and Canada
(24 hours): . . . . . . . . . . . . 949.250.2222
In the UK: . . . 0870 606 2040 - Option 4
In Ireland: . . . . . . . 01 8211012 Option 4
Caution: Federal (USA) law restricts
this device to sale by or on the order of
a physician.
This product is manufactured and sold
under one or more of the following US
patent(s): US Patent Nos. 6,045,512;
and 6,371,923; and corresponding
foreign patents. Additional patents
pending.
Prices, specifications and model
availability are subject to change
without notice.
Refer to the symbol legend at the
end of this document.