9. Specifications
9.3 Permissible temperature
Place, situation
Ambient
Medium
CIP and SIP
EN
Autoclaving
Storage, transport
9.4 Temperature effect
When the temperature of the measuring system deviates from the
reference temperature (+20 °C): max. ±0.5 %/10 K of full scale value
9.5 Ingress protection
IP54 per IEC/EN 60529
9.6 Reference chamber
Material: Stainless steel
Overpressure safety: 75 bar
For further specifications see WIKA data sheet PM 04.17 and the
order documentation.
20
WIKA operating instructions disphragm pressure gauge, model PG43SA-D
Permissible temperature
-20 ... +60 °C
-30 ... +150 °C
150 °C continuously for wetted parts
≤ 134 °C, ≤ 20 minutes
-30 ... +70 °C