Press and release the pedal and inject
the distilled water with the syringe. The
"flush" cycle on the "Polishing" part lasts
about 40 seconds.
Repeat points 26 to 29 another 4 times;
Repeat points 20 to 23 twice until the
irrigation circuit is completely drained of
the water used in the previous steps.
NOTE: It is advisable to drain the circuit
during long periods of non-use. If the
device is used after the disinfection
procedure, and provided different liquids
don't need to be used for irrigation,
the water circuit does not need to be
drained.
At the end of the disinfection procedure,
remove the syringe (Ref. 1), the hose (Ref.
2) and unscrew the male coupling used
(Ref. 3).