• In endodontic root canal therapy, never
operate the files outside of the root
canal to prevent them from breaking.
To prevent breakages, create a smooth
path with a manual endo file and plan
the most direct possible path of access
to minimise any bending of the insert.
5.6
Instructions for Use - Polisher Part
After having connected all the accessories as described in
as follows:
Lift the quick coupling on the air-polishing
handpiece cord. The POLISHING function
is activated.
CAUTION: Make sure the scaler
handpiece is stowed in its housing,
otherwise the system will remain
inactive.
Choose the type of air-polishing
handpiece depending on the operation
that needs to be performed;
CAUTION: The SUBGINGIVAL PERIO
TIP can only be used with the perio air-
polishing handpiece.
On the quick coupling of the cord, slide
the ring and hold it in position, then
insert the air-polishing handpiece until
it comes into contact, and finally release
the ring again;
36
Make delicate movements. Examine the
file often for signs of wear. If the file
breaks inside the canal, avoid contact
between the instrument and the broken
file to stop it from being pushed further
it. Do not apply pressure on the insert in
an axial direction.
Chapter 4.3 on page 18
1
2
3
proceed
refill