Cleaning and Disinfecting the MDS Lite

WARNING:
Caution:
Clean and
disinfect

Cleaning the Peripherals

For information about cleaning the peripherals, see the manufacturer's operating instructions.
20
Cleaning and Disinfecting the MDS Lite
To avoid electrical shock, remove the system and extra battery from the mini-dock
and unplug from power supply before cleaning.
To avoid injury, always use protective eyewear and gloves when performing
cleaning and disinfecting procedures.
To avoid infection, if a pre-mixed disinfectant solution is used, observe the solution
expiration date, and ensure that the date has not passed.
To avoid infection, the level of disinfection required for a product is dictated by the
type of tissue it contacts during use. Ensure the solution strength and duration of
contact are appropriate for the equipment. For information, see the disinfectant
label instructions and the recommendations of the Association for Professionals in
Infection Control and Epidemiology (APIC) and FDA.
Use only recommended cleaners or disinfectants on surfaces. Immersion-type
disinfectants are not tested for use on surfaces.
The exterior surface of the MDS Lite can be cleaned and disinfected using a
recommended cleaner or disinfectant.
1
Remove the system, unplug the power supply, and detach any cables.
2
Clean the surfaces using a soft cloth lightly dampened in a mild soap or
detergent cleaning solution.
Apply the solution to the cloth rather than the surface.
3
Wipe the surfaces with the disinfecting solution. Theracide disinfectant is
recommended.
4
Air dry or towel dry with a clean cloth.