Imaging plate unit
B - 18
CAUSE / SOLUTION
First remove and then reconnect the Ethernet cable
between the unit and PC to see if this solves the
problem.
i.
Physical connection is okay (link active), but the
Dental imaging software animation (fl ashing
yellow) appears on the unit display. The unit is
not confi gured correctly. Reconfi gure.
ii. Cabling not correct (link not active), and the
Unit connection animation appears on the unit
display. The unit is not physically connected to
the PC. Connect the unit to the PC.
iii. The PCs Ethernet connector(s) is faulty. There
is usually a green link LED near the PCs Ether-
net connector(s) on the NIC. If the LED is not on
replace the NIC.
PROBLEM
The unit display is not active but the power on / off
status light is on, green or yellow.
CAUSE / SOLUTION
•
Green LED - defective display or its cabling.
Replace
•
Yellow LED - unit is in standby (energy save)
mode.
B. Installation and setup