Henry Schein MD300K2 Instrucciones De Uso página 97

Pulsooxímetro de mano
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 32
Guidance and manufacturer's declaration- electromagnetic immunity
The model MD300K2 Pulse Oximeter is intended for use in the electromagnetic environment specified
below. The customer or the user of the model MD300K2 Pulse Oximeter should assure that it is used in
such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-4
Electrostatic
transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short in-
terruptions and voltage
variations on power
supply input lines IEC
61000-4-1
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
IEC 60601 test level
±2,4,6,8kV contact
±2,4,8,15kV air
±2kV for power supply
lines ±1kV for input/
output lines
±2kV common mode
<5% UT ( >95%dip
in UT) for 0.5 cycles
40% UT (60%dip in
UT) for 5 cycles 70%
UT( 30%dip in UT) For
2 5 cycles <5% UT
(>95%dip in UT) for
5 sec
30 A/m
50Hz/60Hz
Compliance level
±2,4,6,8kV contact
±2,4,8,15kV air
±2kV for power supply
lines ±1kV for input/
output lines
mode ±2kV
common mode
<5% UT ( >95%dip
in UT) for 0.5 cycles
40% UT (60%dip in
UT) for 5 cycles 70%
UT (30%dip in UT)
For2 5 cycles <5%
UT ( >95%dip in UT)
for 5 sec
30 A/m
50Hz/60Hz
Electromagnetic
environment-guidane
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
model MD300K2 Pulse
Oximeter requires
continued operation
during power main
Interruptions, it is
recommended that the
model MD300K2 Pulse
Oximeter be powered
from an uninterruptible
power supply or a
battery.
Power frequency ma-
gnetic fields should be
at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
97
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

900-8275900-8277900-8278

Tabla de contenido