Maintenance And Repair; Maintenance - Henry Schein MD300K2 Instrucciones De Uso

Pulsooxímetro de mano
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 32
Warnings!
When the alarm occurs, check patients' conditions immediately.
• When the alarm occurs, check patients' conditions immediately.
• Check which parameter is alarming or which alarm is going on.
• Check patient's condition.
• Search for the source of alarm.
• Make the alarm mute if necessary.
• Check the alarm when no warning.
Alarm delay
The alarm condition delay and alarm signal generation delay: less than 1s.
Note:
1. The pulse rate correspondence with the user's pulse rate. It based on the user's actual pulse rate.
2. Use the alarm setting in di erent areas will be brings the potential dangerous.
3. The alarm setting can recover if the power-break time less than 30s.
4. Set the high parameter value with simulator to test the e ciency of the alarm system.
5. Do not set the parameter value out of the range, or the alarm system will failure.
6. The device can reserve alarm setting if it power break.

8 MAINTENANCE AND REPAIR

Warnings!
• The advanced circuit inside the oximeter does not require periodic calibration and maintenance, except
replacing the batteries.
• Don't open the cover of oximeter or repair electronic circuits. Its open will cause the damage of the device
and the annulment of the guarantee.

8.1 Maintenance

Use only the substances approved by us and methods listed in this chapter to clean or disinfect your equip-
ment. Warranty does not cover damage caused by unapproved substances or methods.
We make no claims regarding the e cacy of the listed chemicals or methods as a means for controlling
infection. For the method to control infection, consult your hospital's Infection Control O cer or Epidemiolo-
gist. Keep your equipment and accessories free of dust and dirt. To avoid damage to the equipment, follow
these rules:
• Always dilute according the manufacturer's instructions or use lowest possible concentration.
• Do not immerse part of the equipment into liquid.