7 ALARM
ALARM PRIORITY:
There are two-level priorities for selection.
High priority: the highest level alarm, indicates the patient is in the very dangerous situation.
Low priority: indicate the technical alarm caused by the device itself.
Alarms of the oximeter include technical and physiological alarms. All the two priorities
divided by built-in module and cannot be changed by user.
Assignment of priority:
Paramter
Value
Alarm lamp
Lamp Frequency
Audiblesound
Alarm cycle
Alarm info
Notes:
1. The alarm will appear if the measurement value out of range.
2. The alarm sound will go on until alarm disappears or is turned o .
3. After silencing the alarm, the corresponding indicator will indicate this.
4. The power low alarm: the corresponding indication lamp will be flashing with a red frame.
High
SpO /PR
Red
Flashing
1.5Hz
Di- Di – Di ----- Di - Di
3 s
SpO too high/low, PR too high/low, Battery
power low
Fig.6.5
Low
/
/
/
/
Di
20 s
Probe o /Finger o