aDVancED mEnu controlS
Advanced Menu List
Display Example
Description
This is the default screen when the SleepCube is turned on. The actual mask pressure and heater setting are shown.
P: 5.0
CPAP
The displayed pressure may vary slightly while breathing. If the heater system is not connected, the heat setting will
Heat:Off
be blank.
This is a prescription setting which controls the operating pressure.
Set Pressure
10.0 cmH20
Selectable delay time setting. Your provider has set the delay time. If the unlocked symbol is shown, you are able to
Delay Time
adjust this setting. If the locked symbol is shown, this setting can only be adjusted by your provider.
45 minutes
This is a prescription setting which controls the operating pressure during the delay time.
Delay Pressure
5.0 cmH20
Pressing the down key enters the Enable menu. Refer to next section for Enable Menu details.
Enable Menu
Press
This setting changes the contrast for the LCD display. This only needs to be adjusted if you find it difficult to see the
LCD Contrast Adj
display from your viewing angle.
50 %
The total number of hours (in 1/10 hour increments) that the SleepCube has been providing therapy.
Compliance Meter
123456.7 hours
The total number of hours (in 1/10 hour increments) that the SleepCube has been in operation.
Hourmeter
123456.7 hours
Firmware Version
Shows version of installed firmware.
V0.01 01/06/2006
Serial Number
Shows the serial number of the device.
HD123456
Shows the maximum adherence achieved for any period of thirty consecutive days of use within a ninety-day time
Adherence Score
frame. The score represents the percentage of days, within thirty consecutive days, where usage met or exceeded
83% (25/30)
the 'Usage Threshold.'
SmartCode
SmartCode 1-Day
numbers, and includes a special character for error checking to ensure accuracy. Use the UP arrow or DOWN
G201-0Y7-6GPH
arrow keys to change the SmartCode reporting period.
Enable Menu List
The Enable menu is available in all modes and is a sub-menu that allows you to control various functions of your SleepCube. It is accessed by
pressing the down button while in this Advanced Menu.
Display Example
Description
This option controls the Auto-ON feature, which automatically starts the flow of air after breathing once or twice
Auto-ON
into the mask.
Enabled
This option controls the Auto-OFF feature, which automatically stops the flow of air after taking off the mask. The
Auto-OFF
device may require up to 30 seconds to turn OFF depending on the kind of mask used.
Enabled
Exit Menu
Pressing the up key exits the Enable menu.
Press
is used for therapy management reporting; therapy information is encoded in a series of letters and
TM
EN
SE-DV51