Immunity test
Electrostatic
+/- 6 kV contact
discharge
+/- 8 kV air
(ESD)
IEC 61000-4-2
Electrical fast
+/- 2 kV
transient / burst
for power supply unit
IEC 61000-4-4
+/- 1 kV
for input/output lines
Surge
+/- 1 kV
IEC 61000-4-5
differential mode
+/- 2 kV
common mode
Voltage
dips,
<5% U
short
(>95% dip in U
interruptions and
For 0,5 cycle
voltage variations
on power supply
40% of U
input lines
(60% dip in U
IEC 61000-4-11
For 5 cycles
70% of U
(30% dip in U
For 25 cycles
<5% U
(>95% dip in U
For 5 sec
Power frequency
3 A/m
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. main voltage prior to application of the test level.
T
Operation and
maintenance manual
IEC 60601-1-2
test level
+/- 6 kV contact
+/- 8 kV air
+/- 2 kV
for power supply
lines
+/- 1 kV
for input/output lines
+/- 1 kV
differential mode
+/- 2 kV
common mode
<5% U
T
)
(>95% dip in U
T
For 0,5 cycle
40% of U
T
)
(60% dip in U
T
For 5 cycles
70% of U
T
)
(30% dip in U
T
For 25 cycles
<5% U
T
)
(>95% dip in U
T
For 5 sec
3 A/m
MU_025
Rev.2
Compliance level
T
)
T
T
)
T
T
)
T
T
)
T
04/02/16
Page 24 of 28
Electromagnetic
environment-guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of the
Product requires continued
operation during power mains
interruptions, it is
recommended that the
Product be powered from an
uninterruptible power supply
or battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
EN