Electromagnetic Compatibility Information

Table 1
Guidance and manufacturer's declaration – electromagnetic emissions
The BM 19 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 19 should assure
that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000­3­2
Voltage fluctuations/flicker emissions
IEC 61000­3­3
Table 2
Guidance and manufacturer's declaration – electromagnetic immunity
The BM 19 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 19 should assure
that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000­4­2
Power frequency (50/60 Hz)
magnetic field IEC 61000­4­8
All manuals and user guides at all-guides.com
Compliance
Electromagnetic environment – guidance
Group 1
The BM 19 uses RF energy only for its internal function. Therefore, its RF emissi­
ons are very low and are not likely to cause any interference in nearby electronic
equipment.
Class B
The BM 19 is suitable for use in all establishments, including domestic establish­
ments and those directly connected to the public low­voltage power supply net­
Not applicable
work that supplies buildings used for domestic purposes.
Not applicable
IEC 60601 test level Compliance level Electromagnetic environment – guidance
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30 %.
Power frequency magnetic fields should be at levels characte­
ristic of a typical location in a typical commercial or hospital
environment.