Electromagnetic Compatibility Information
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration – electromagnetic emissions
The BC 58 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BC 58 should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The BC 58 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 58 should assure
that it is used in such an environment.
IMMUNITY test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
All manuals and user guides at all-guides.com
Compliance
Electromagnetic environment – guidance
Group 1
The BC 58 uses RF energy only for its internal function. Therefore, its RF emissi-
ons are very low and are not likely to cause any interference in nearby electronic
equipment.
Class B
The BC 58 is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
Not applicable
supplies buildings used for domestic purposes.
Not applicable
IEC 60601
Compliance
test level
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30 %.
Power frequency magnetic fields should be at levels characteristic of
a typical location in a typical commercial or hospital environment.
106