Instructions For Use - Bego VarseoSmile Crown plus Instrucciones De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 8
VarseoSmile Crown
en

Instructions for use

VarseoSmile Crown
plus
Resin for 3D printing of permanent single crowns, inlays, onlays and
veneers.
1. Intended use / Indication
VarseoSmile Crown
is a light-curing, free-flowing plastic based on
plus
methacrylic acid esters for the production of permanent single crowns, in-
lays, onlays and veneers.
2. Contraindications
Known allergy to one or more ingredients. In cases of doubt, the allergy should
be clarified and ruled out based on a specific test prior to the application
of this product. VarseoSmile Crown
should not be used for purposes other
plus
than the production of permanent single crowns, inlays, onlays and veneers.
Any deviation from these instructions for use can have negative effects on the
chemical and physical quality of plastics made from VarseoSmile Crown
3. Safety instructions
This product is produced and tested according to the most stringent quality
standards. It may only be used by a dental professional. In order to ensure
optimum further processing, please read the information contained in the
instructions for use carefully.
The safety and precautions included in these instructions for use and safety
data sheet shall apply to the handling of liquid resin and printed objects that
have not been post-cured (objects in the "green condition").
0197
4. Side effects and precautions
Precautions / Protection
It is essential that protective clothing be worn when handling this product.
Safety goggles and nitrile gloves must be used. Further information on hand-
ling the product can be found in the safety data sheet and also downloaded
from the BEGO Download Centre at www.bego.com. We cannot completly rule
out adverse reactions (e. g. intolerance or allergies) to specific material com-
ponents for all individuals. In such isolated cases, the user should discon-
tinue use of the material.
Hazard statements as per MSDS
• Causes skin irritation.
• May cause an allergic skin reaction.
• Causes serious eye irritation.
• May cause respiratory irritation.
WARNING
• May cause long harmful effects to aquatic life.
Precautionary statements as per MSDS
Contains:
• Avoid breathing mist / vapours / spray.
Esterification products of
• Avoid release to the environment.
4,4'-isopropylidiphenol,
• Wear protective gloves / protective clothing / eye
ethoxylated and 2-met-
protection / face protection.
hylprop-2enoic acid.
• Call a POISON CENTER / doctor if you feel unwell.
Silanized dental glass,
• If skin irritation or rash occurs: Get medical
methyl benzoylformate,
advice / attention.
diphenyl(2,4,6-trimethyl-
• If eye irritation persists: Get medical advice /
benzoyl) phosphine oxide.
attention.
Total content of inorganic
• Dispose of contents / container as per local and
fillers (particle size 0.7 μm)
national regulations.
is 30–50% by mass.
Notice: Any serious incident that has occurred in relation to the device due
to a malfunction should be reported to the manufacturer and the competent
authority of the Member State in which the user and / or patient is esta-
blished.
5. General information on handling
Delivery
VarseoSmile Crown
is supplied in seven colours according to the VITA*
plus
classical shade system, in lightproof and sealed bottles.
Filling quatity:
• REF 41107 = 500 g, A1 Dentin
• REF 41120 = 250 g, B1 Dentin
• REF 41117 = 250 g, A1 Dentin
• REF 41111 = 500 g, B3 Dentin
• REF 41108 = 500 g, A2 Dentin
• REF 41121 = 250 g, B3 Dentin
• REF 41118 = 250 g, A2 Dentin
• REF 41112 = 500 g, C2 Dentin
• REF 41109 = 500 g, A3 Dentin
• REF 41122 = 250 g, C2 Dentin
• REF 41119 = 250 g, A3 Dentin
• REF 41113 = 500 g, D3 Dentin
• REF 41110 = 500 g, B1 Dentin
• REF 41123 = 250 g, D3 Dentin
Storage
This product must be stored in the original sealed bottle, or in the car-
tridge at room temperature (approx. 22 °C) in a dark, dry place. It must
be ensured that the temperature does not drop below +4 °C and does not
exceed +28 °C! The minimum shelf life date printed on the product must
be observed.
Notice: Expected results cannot be guaranteed if materials which have
exceeded their minimum shelf life date are used or if storage instructions
are not followed.
The completely cured print objects must be stored at room temperature and
protected from sources of light.
* This symbol is a commercial designation / registered trademark of a
company that is not part of the BEGO company group.
plus
6. Processing requirements
1. Design
• Create the object (STL-file) using a commercial CAD software, which is
intended for dental applications.
• When designing, observe the requirements for minimum wall thick-
nesses for finished restorations.
Notice:
Crowns, inlays, onlays and veneers
Minimum wall thicknesses anterior teeth
Minimum wall thicknesses posterior teeth
Minimal wall thickness, cervical
Minimal wall thickness, veneer shell
2. Nesting & preparation for printing
plus
.
• Import STL file
• Manual / Automatic rotation and placement
• Optimal orientation: horizontal, occlusal plane facing the build
platform
• Manual / Automatic generation of supports
3. Printing
VarseoSmile Crown
plus
has been verified and validated in combination
with various system components (3D printers, cleaning devices and
post-curing devices). We are constantly working on further qualifications.
You can find these compatible system components on our website
https://www.bego.com/3d-printing/compatibility-overview/
Please pay special attention to the build platform and resin tank mate-
rials as noted in the compatibility matrix.
An example list of compatible 3D additive manufacturing printers and
their operation software:
Compatible 3D Printers
3D Printer Model
Printer Firmware
Varseo
1.14 or higher
Varseo L
1.02 or higher
Varseo S
1.14 or higher
Varseo XS
2.6.8.24 or higher
ASIGA MAX UV
2018-09-03
SprintRay Pro 55
6.32.5
SprintRay Pro 95
2.5.1
WARNING: This material is suitable for manufacturing highly
reliable dental products only when using BEGO approved com-
patible systems including the material parameters. If unapproved
components or material parameters are used, there is a high risk
of unreliable and / or unusable products which may endanger the
safety of the user.
Notice: It is important to follow the instructions for use and maintenance
instructions provi ded by the manufacturer for all system components.
4. Necessary tools, equipment and materials for post-processing
• Stainless steel spatula
• Unheated ultrasonic bath
• Ethanol solution 96 %
• Spray bottle with 96 % ethanol solution
• Cutting wheel or side cutters (for support structure removal)
• Sandblaster 1.5 bar
• Glass bead blasting material 50 μm (e. g. Perlablast
REF 46092 / 54302)
7. Processing
The following instructions contain details of a validated workflow for the
3D printing process with a compatible 3D printer.
The ideal working temperature of VarseoSmile Crown
ture range between 18 °C and 28 °C. Before use, the resin must be homoge-
neous. Before the first use, the material has to be shaken well about 2 min.
When decanting, make sure that the printing resin is exposed to daylight for
as short a period of time as possible. Mix the resin in the cartridge / resin
tank if a transparent layer is visible on the surface.
For further processing – selecting the resin, material parameter, setting up
the print job – as part of the printing process, follow the respective printer
instructions for use.
Cleaning and preparation for post-curing
On completion of printing, the print objects are released from the build
platform using the spatula. The print object should be cleaned in two steps
with ethanol (96 %) using an ultrasonic bath.
Precaution: Never fill ethanol directly into the ultrasonic bath; place
it in the recommended container (REF 19621) in the ultrasonic bath
filled with water. Use an explosion-proof ultrasonic bath.
1. Clean the print object for 3 min in a reusable ethanol solution (96 %)
using an unheated ultrasonic bath.
2. The precleaned object must be cleaned thoroughly for 2  min using a
fresh ethanol (96 %) solution with the aid of an unheated ultrasonic
bath. The print object is then removed from the ethanol bath and sprayed
with additional ethanol (96 %) in order to fully rinse off any remaining
resin residue.
Tip: Resin residues can also be removed using a brush soaked in ethanol
(96 %).
After cleaning, the print object is dried using compressed air under an
extraction unit. If there is liquid resin still adhering to the surface of the
object, this can be completely removed by spraying again with ethanol
(96 %) and re-drying.
Preparation for post-curing
• Remove the support structure with the help of a cutting wheel or side
1.0 mm
1.0 mm
• Remove the white layer using a glass bead blasting material 50 μm
1.0 mm
0.5 mm
• Check for fit and finish the objects completely. Finishing and countouring
Post-curing process
The final properties of the printed object depend on the post-curing process.
Please note the assignment of the light curing device to the 3D printer of the
approved system components.
The post-curing of the object is done without use of a model, then allow to
cool for 3–5 minutes or until the object feels cool.
VarseoSmile Crown
various system components (3D printers, cleaning devices and post-curing
devices). You can find these compatible system components on our website
https://www.bego.com/3d-printing/compatibility-overview/
An example list of compatible post-curing devices:
Post-curing
3D Printer
BEGO Varseo
BEGO Varseo L
Nesting Software
Provider
BEGO Varseo S
BEGO
BEGO Varseo XS
CAMCreator Print
BEGO
ASIGA MAX UV
Version 1.14
SprintRay Pro 55 SprintRay ProCure 2 x 20 min at 20 °C
or higher
SprintRay Pro 95 SprintRay ProCure 2 x 20 min at 20 °C
ASIGA Composer
ASIGA
RayWare
SprintRay
RayWare
SprintRay
Note: When employing the BEGO Otoflash, use the protective gas function.
This results in a further reduction of the already low remaining monomer
content. To do so, set the protective gas function to switch position  1.
Details can be found in the instruction manual for the post-curing device.
It is also important to follow maintenance instructions provided by the
manufacturer for all system components.
The times given only apply to regularly maintained equipment that guaran-
tees a correspon ding light intensity.
micro,
®
Supplementing / Repairing of printed objects
Defects (e. g. missing contact points, fractures, and so on) can be
supplemented with the resin or commercially available composite
veneering materials.
is in the tempera-
plus
Supplementing with resin
• Blast the areas to be supplemented with aluminum oxide (e. g. Korox
• Put some liquid VarseoSmile Crown
• Put it under light for a short time (e. g. 5 flashes) in the BEGO Otoflash.
• If more material has to be applied, put additional liquid Varseo Smile
• Follow the instructions in the section „post-curing process" for final
Supplementing with veneering material
The printed objects can be supplemented with commercially available com-
posite veneering materials (e. g. VITAVM LC*, Vita Zahnfabrik, and VITAVM
LC flow*, Vita Zahnfabrik). The manufacturer's instructions for use must
be observed.
Polish
Polish the surface of the objects with pumice stone and polishing com-
pound. Avoid overhea ting of the restorations during polishing. Optimal
surface quality is achieved by polishing after post-curing.
Tip: Optionally, the surface of the objects can be coated with light-curing
glaze (e. g. Vita ENAMIC GLAZE *, Vita Zahnfabrik or GC OPTIGLAZE *, GC).
Pay attention to the manufacturer`s instructions for use.
4
Precaution: The entire cleaning process should not take longer than
5 minutes as this could otherwise have a detrimental effect on the
printed objects (swelling of the object with ethanol).
cutters.
(e. g. Perlablast
micro, REF 46092 / 54302) at a maximum blasting
®
pressure of 1.5 bar.
can be performed using carbide cutter or diamond grinding stones.
plus
has been verified and validated in combination with
Light-curing
Exposure Cycles
Device
BEGO-Otoflash
2 x 1,500 flashes
HiLite-Power*
2 x 90 seconds
WARNING: This material is suitable for manufacturing highly reli-
able dental products only when using BEGO approved compatible
systems including the material parameters. If unapproved compo-
nents or material parameters are used, there is a high risk of unre-
liable and / or unusable products which may endanger the safety of
the user.
Precaution: If the post-curing process is interrupted by a malfunc-
tion, the printed object should not be used until it has been cured
for one full cycle.
Refer to the operating instructions for the light-curing device to find
out how the error can be resolved, and then repeat the post-curing
process with the printed objects.
110, BEGO) – pressure 1.5 bar and particle size 110 µm.
plus
on the object.
Crown
on the last layer and then put it again under light for a short
plus
time (e. g. 5 flashes) in the BEGO Otoflash.
curing of the objects.
Caution: The dental object may only be repaired or supplemented
outside the patient's mouth and by a dental professional.
Additional
Information
Turn object between
the exposure cycles
®
Tabla de contenido
loading

Tabla de contenido