Ambient Conditions - Dräger Alcotest 6000 Instrucciones De Uso

Ocultar thumbs Ver también para Alcotest 6000:
Tabla de contenido
Measurement units
mg/L, μg/100 mL, μg%, μg/L, g/210 L, g/L ‰, %, mg/100 mL, mg%, mg/mL
The instrument meets the requirements of EN 15964. It can be configured
according to the standard.
Measured value display with configuration according to EN 15964:
Measured value
0.0 mg/L to 0.03 mg/L
<0.03 mg/L

Ambient conditions

During operation
During storage
Sampling
Respiratory volume
Breath flow
Breath length
Measurement accuracy
Standard deviation of repeatability with ethanol standard
Up to 0.50 mg/L
≥0.50 mg/L
Up to 1.00 ‰
≥1.00 ‰
Drift of sensitivity
Display
0.0 mg/L
Measured value
–5 to +50 °C (23 °F to +122 °F)
10 to 100 % RH (non-condensing)
600 to 1300 hPa
–20 to +60 °C
(–4 °F to +140 °F)
15 to 75 % RH (10 to 100 % RH <48 h)
>1.2 L (default setting)
>10 L/min (default setting)
>1.6 s (default setting)
0.008 mg/L
1.7 % of the measured value
0.017 ‰ or
1.7 % of the measured value
Depending on which value is higher
Typically 0.4 % of the measured value per
month
Technical data
en
29
Tabla de contenido
loading

Tabla de contenido