Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 118
Table of contents
Installation
on page 9

Specifications

Specifications are subject to change without notice.
Specification
Details
Dimensions (W x D x H)
160.5 x 432.2 x 297 mm (6.32 x 16.66 x
11.7 in.); with SVS, D=287 mm (15.2 in.)
Weight
4.8 kg (10.5 lb)
Enclosure
IP68 waterproof rating, polystyrene
Pollution degree
3
Protection class
III
Installation category
I
Operating temperature
-10 to 50 °C (14 to 122 °F)
Storage temperature
-40 to 60 °C (-40 to 140 °F)
Altitude
4000 m (13,123 ft) maximum
Power requirements
Supplied by FL900 Series Logger, Flo-Logger
Interconnecting cable
Polyurethane, 0.400 (±0.015) inch diameter
(disconnect at both sensor and
logger ends)
Standard length: 9 m (30 ft); maximum length:
305 m (1000 ft)
All manuals and user guides at all-guides.com
Operation
on page 20
Specification
Depth measurement
Method: Ultrasonic
Standard operating range from Flo-Dar sensor
housing to liquid: 0-152.4 cm (0-60 in.)
Optional extended operating range from
transducer face to liquid: 0-6.1 m (0-20 ft)
(with 43.18 cm (17 in.) deadband), temperature
compensated
Accuracy: ±1%; ±0.25 cm (±0.1 in.)
Surcharge depth measurement
Method: Piezo resistive pressure transducer
with stainless steel diaphragm
Auto zero function maintains zero error <
0.5 cm (0.2 in.)
Range: 3.5 m (138 in.); overpressure rating:
2.5 × full scale
Velocity measurement
Method: Radar
Range: 0.23-6.10 m/s (0.75-20 ft/s)
Frequency Range: 24.075 to 24.175 GHz,
15 mW (EIRP) maximum
Accuracy: ±0.5%; ±0.03 m/s (±0.1 ft/s)
Certifications
The Flo-Dar transmitter is certified to the
requirements that follow:
• Transmitter type: Field disturbance sensor
• Frequency: 24.125 GHz - Doppler pulse
• Maximum rated power output: 128 dbuV
Certified to:
FCC Part 15.245: FCC ID: VIC-FLODAR24
(average) at 3 m (9.8 ft)
English 3