Function "Video output signal selection":
Rotate the encoder (5) to select icon
confirm.
Rotate the encoder (5) to select the required video output signal (PAL or
NTSC).
Press the encoder shortly to confirm.
To exit the menu, keep the encoder (5) pressed for two seconds or wait
10 seconds to enable automatic exit.
10
USE OF THE VIDEO OUT JACK
"Video out" jack (11) is designed to connect external recording devices
and to transmit video signal to monitors, TV sets etc.
Select a suitable video output standard - PAL or NTSC (refer to section
9 "MENU").
Use the included video cable to connect a peripheral device to the
"Video out" jack (11).
Switch on your thermal imaging scope – a peripheral device will show
the image seen through the scope.
11
ACCESSORIES
Using the Weaver mounting rail (8), you can attach various accessories to
the Quantum scopes, such as:
Battery packs EPS3/EPS5 (SKU#79111/79112)
CVR640 video recorder (SKU#17044) and others.
12

CARE AND MAINTENANCE

The thermal imaging scopes have IPX4 degree of protection (protected
against splash water) but they are not intended for submersion into
water.
Attempts to disassemble or repair the scope will void the warranty!
8
and press the encoder to
Clean the scope's optical surfaces only if necessary, and use caution.
First, remove (by blowing with a blower brush or canned air) any dust
or sand particles. Then proceed to clean by using camera/lens
cleaning equipment approved for use with multicoated lenses. Do not
pour the solution directly onto the lens!
The scope can be used in operating temperatures ranging from
-20 °C to +50 °C. However, if it has been brought indoors from cold
temperatures, do not turn it on for three to four hours. This will prevent
external optical surfaces from fogging.
Always store the unit in its carrying case in a dry, well-ventilated space.
Remove the batteries if the scope is not going to be used for an
extended period (longer than one month).
Batteries shall not be exposed to excessive heat such as sunshine, fire
or the like.
13
TROUBLESHOOTING
Listed below are some potential problems that may occur when using the
scope. Carry out the recommended checks and troubleshooting steps in
the order listed. Please note that the table does not list all of the possible
problems. If the problem experienced with the scope is not listed, or if the
suggested action meant to correct it does not resolve the problem,
please contact the manufacturer.
problem
possible cause
The unit will
Batteries have been wrongly
not turn on.
installed.
Oxidized contact points in the
battery compartment due to
"leaky" batteries or contact
points becoming exposed to
a chemically reactive solution.
The batteries are fully
exhausted.
The image is blurry, with
Calibration is to be done.
vertical stripes and uneven
background.
The image is too dark.
The lowest brightness level
is set.
corrective action
Reinstall the batteries observing polarity.
Clean the battery compartment,
focusing on the contacts.
Install fresh batteries.
Do the calibration according to section
8 "OPERATION"
Adjust brightness by rotating the
encoder (5).
9