3.5 Probes

3.5.1 Waterproof Probes

2.0MHz/3.0MHz waterproof probe can be connected with the main unit of the SONOTRAX
Series.
Figure 3-7 2.0MHz waterproof probe
The main information on the probe is as follows:
CD2.0/3.0: CD means continuous wave Doppler, 2.0/3.0 means central frequency is 2.0
MHz/3.0MHz.
MS3-14320: Part number of the 2.0MHz waterproof probe.
MS3-14321: Part number of the 3.0MHz waterproof probe.
A: Version number of the probe.
SNxxxxx: Serial number of the probe.
Waterproof: The probe is waterproof.
IPX8: Water Ingress Protection Code, which indicates this probe can work continuously for 5
hours when being immersed into water within 1 meter depth.

3.5.2 Waterproof Vascular Probes

The 4.0MHz (optional) and the 8.0MHz waterproof vascular probes are used for monitoring
arteries and veins.
Figure 3-9 4.0MHz Waterproof Vascular Probes
Figure 3-8 3.0MHz waterproof probe
Figure 3-10 8.0MHz Waterproof Vascular Probes
- 10 -
SONOTRAX Series Ultrasonic Pocket Doppler User Manual
The main information on the probe is as follows.
CD 4.0/CD8.0: CD means continuous wave Doppler, 4.0/8.0 means the central frequency is
4.0/8.0 MHz.
MS3-14346: EDAN part number of 4.0MHz waterproof vascular probe.
MS3-14347: EDAN part number of 8.0MHz waterproof vascular probe.
A: the version number of this vascular probe.
SNxxxxx: Serial number of this vascular probe.
Waterproof: The probe is waterproof.
IPX8: Water Ingress Protection Code, which indicates this probe can work continuously for 5
hours when being immersed into water within 1 meter depth.

3.5.3 Probe Socket

Figure 3-11 Probe socket
Jack
1
2
3
4
5
6
The probe socket is shown as Figure 3-11.
Connect the 2.0MHz waterproof probe, 3.0MHz waterproof probe, 4.0MHz waterproof
vascular probe or 8.0MHz waterproof vascular probe developed by EDAN to the
SONOTRAX Series according to real requirements.
WARNING
: Do not try to connect any other plug to the probe socket
except the plug of the probes mentioned above.
WARNING
: In case of disengaging from the probe interface, the
connecting cable of the probe can not be dragged for more
than 2 meters.
Definition
Power Supply
Signal
Probe Coding 1
Probe Coding 2
Probe Coding 3
(Shell) GND