EDAN INSTRUMENTS SONOTRAX Serie Manual Del Usuario página 45

Manuel Utilisateur de Série de Poche Ultrasonique Doppler SONOTRAX
Appendix
A
EMC
Information-Guidance
Manufacture's Declaration
A1.1 Electromagnetic Emissions - for all Equipment and
Systems
Guidance and manufacture's declaration–electromagnetic emission
The SONOTRAX Series Ultrasonic Pocket Doppler is intended for use in the electromagnetic
environment specified below. The customer or the user of the device should assure that it is used in such
an environment.
Emission test
Compliance
The SONOTRAX Series Ultrasonic Pocket Doppler uses RF
RF emissions
energy only for its internal function. Therefore, its RF emissions
Group 1
CISPR 11
are very low and are not likely to cause any interference in nearby
electronic equipment.
The SONOTRAX Series Ultrasonic Pocket Doppler is suitable
for use in all establishments, including domestic establishments
RF emission
Class B
and those directly connected to the public low-voltage power
CISPR 11
supply network that supplies buildings used for domestic
purposes.
A1.2 Electromagnetic Immunity - for all Equipment and
Systems
idance and manufacture's declaration–electromagnetic immunity
Gu
The SONOTRAX Series Ultrasonic Pocket Doppler is intended for use in the electromagnetic
environment specified below. The customer or the user of the device should assure that it is used in such
an environment.
IEC 60601 test
Immunity test
level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
and
Electromagnetic environment–guidance
Electromagnetic
Compliance level
environment-guidance
Floors
should
be
concrete or ceramic tile. If
±6 kV contact
floor
are
covered
±8 kV air
synthetic material, the relative
humidity should be at least
30%.
- 25 -
A1.3 Electromagnetic Immunity - for all Equipment and
Systems that are not Life-supporting
The SONOTRAX Series Ultrasonic Pocket Doppler is intended for use in the electromagnetic
environment specified below. The customer or the user of the device should assure that it is used in such
an environment.
Immunity test
Radiated RF
IEC 61000-4-3
NOTE 1
NOTE 2
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
wood,
the location in which the SONOTRAX Series Ultrasonic Pocket Doppler is used exceeds the
applicable RF compliance level above, the SONOTRAX Series Ultrasonic Pocket Doppler should be
with
observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the SONOTRAX Series Ultrasonic Pocket Doppler.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Manuel Utilisateur de Série de Poche Ultrasonique Doppler SONOTRAX
Guidance and manufacture's declaration – electromagnetic immunity
Compliance
IEC 60601 test level
level
3 V/m
80 MHz to 2.5 GHz
3 V/m
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
- 26 -
Electromagnetic environment-guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the SONOTRAX Series Ultrasonic
Pocket Doppler, including cables, than the
recommended
separation
distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
3
5 .
=
d
P
80 MHz to 800 MHz
E
1
7
=
d
P
800 MHz to 2.5 GHz
E
1
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a
survey,
should
be
less
than
the
compliance level in each frequency
b
range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:

Hide quick links:

loading

Productos relacionados para EDAN INSTRUMENTS SONOTRAX Serie