6.4.1. "ALARMS" list
Below is the list of "ALARMS"
ALARM
ENGINE BLOCK
ALARM
SHORT CIRCUIT
PROBE
PROBE INTERRUPTED
OR DISCONNECTED
CHAMBER OVER
HEATING
OVER HEATING
REMOTE CONTROL
SWITCH ALARM
Page 140
RESTORE
When you see the alarm "engine block" make sure that the
emergency mushroom is unlocked.
Turn off the oven and repeat the operation.
If the alarm is raised again, call the technical support service.
It reports that the chamber temperature probe is in short-
circuiting.
The oven is turned off.
Call the technical support service for the probe replacement.
It reports that the chamber temperature probe is interrupted or
disconnected.
Il oven is turned off.
Call the technical support service for the control and/or
replacement of the probe.
It reports that the temperature in the chamber has exceeded
500°C.
The heating of the oven is turned off.
Turn off the oven, wait for the chamber to cool down
and repeat the operation.
If the alarm is raised again, call the technical support service.
TQ It reports that the temperature of the power card heatsink
has exceeded 70°C.
Turn off the oven and check that the cooling fans of the
technical compartment are working properly.
TS Indicates that the security thermostat intervened.
Call Technical Support.
Indicates that the general remote control switch is faulty.
Call Technical Support.