Chapter 11 Appendix; Prompt Information Explanations - Gima 35132 Manual De Uso

Ocultar thumbs Ver también para 35132:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

11.1 Prompt information explanations

Mute C-D: XXX seconds
NIBP C-D: XXX seconds
TOUR C-D: XXX seconds
Probe off
PR over limit
SpO
over limit
2
SYS over limit
DIA over limit
MAP over limit
NIBP error 1#
NIBP error 2#
NIBP error 3#
NIBP error 4#
Cuff error
NIBP error 5#
Air leak
NIBP over range
Over motion
Over pressure
NIBP timeout
User Manual for Vital Signs Monitor

Chapter 11 Appendix

Alarm silence count down: XXX seconds
NIBP auto measuring cycle count down: XXX seconds
Hemostat alert count down: XXX seconds
SpO
probe fells off
2
PR value exceeds the high/low alarm limit
SpO
value exceeds the high/low alarm limit
2
Systolic pressure value exceeds the high/low alarm limit
Diastolic pressure value exceeds the high/low alarm limit
MAP value exceeds the high/low alarm limit
Sensor or other hardware error
Very weak signal because of the cuff, or the patient has very weak pulse
Blood pressure amplifier overflow due to excessive movement
Leaking during the pneumatic device testing
Cuff is not wrapped correctly, or is not connected
Abnormal condition of CPU, such as register overflow, divided by zero
Air moving part, tube or the cuff leak air
The measurement range exceeds 255mmHg (for neonates: over 135 mmHg)
The repeated measurement due to moving, excessive noise during the
stepping inflation and measuring pressure and pulse, e.g. during patient
shaking motion
Cuff press exceeds the safety limit value of software. Limit value for adult:
290mmHg; Limit value for pediatric: 145mmHg;
Or caused by cuff extrusion or flapping cuff with force.
Adult measurement is more than 120 seconds, neonate measurement is
more than 90 seconds.
66
Tabla de contenido
loading

Este manual también es adecuado para:

Pc-900plus snet

Tabla de contenido