4.9.5 Patient Info
Screen description:
"Patient ID": change or set current patient's ID number, 0~100 adjustable; Once the patient ID is
²
changed, the history data in trend graph will be cleared, and the parameter settings will be resumed to
default value.
"Category": change or set the category of current patient; three options "Adult" ,"Pediatric" and
²
"Neonate", press OK key to confirm the setting and the patient type indicator will be ON at the front
panel of the device. The default is "Adult".
4.9.6 Nurse Call Setup
Screen Description:
² "Output level": two options "Low" or "High" output levels are available.
² "Duration": two options "Pulse" and "Continuous" output modes are available; Refer to the following
form for the Output level and Duration.
User Manual for Vital Signs Monitor
Figure 4.17 Patient Info. setup screen
Figure 4.18 Nurse Call Setup Screen
42