Audible Alarm Indication
The audible alarm has different tone pitch and on-off beep patterns for each priority category. These are
summarized in the table below.
Alarm Category
High priority alarm
Medium priority alarm
Low priority alarm
Note: Visual alarm indication can not be suspended or removed. Audible alarms may be decreased in volume or
silenced as described.

5.3 Alarm Reset and Silence

Press
(Alarm Silence) key to pause the audible alarm temporarily or reset the current alarm condition.
During the monitoring process, press "Alarm Silence" key shortly to start the alarm silence for 2 minutes. The
counting down time shows up on the upper left corner of the screen once the alarm silence is activated. Long
pressing "Alarm Silence" key will reset the current alarm condition, that means the sound mute for this alarm will
not resume unless another alarm condition occurs. During the alarm silence period, if there is a new alarm
condition other than the current one occurs, the device will generate the audible alarm indication again
automatically. After the end of alarm silence period, if the current alarm condition still exists, then the audible
alarm indication will resume as well.
When the monitor generates alarms, the user can press
for a given silence period when necessary..
M DO NOT silence the audible alarm or decrease its volume if patient safety could be
compromised.
M For the alarm conditions of "Can not detect SpO
alarm indication will only last for about 7 seconds.
M Alarm signal can be reset, but it can NOT be deactivated all the time.

5.4 Alarm Settings

1、 Except volume of audible alarm can be adjustable, the other properties of the alarm signal cannot be
adjusted by the user, such as alarm priority setting, alarm light flashing and so on. In addition, all
alarms in this patient monitor are "non-latched" type, that is to say, when the alarm condition does
not exist, the corresponding alarm signal will automatically stop.
The alarm volume range is shown as below:
High: 45dB~80dB (The distance from the front of device to the test instrument is 1m)
²
Medium: 45dB~75dB (The distance from the front of device to the test instrument is 1m)
²
Low: 45dB~70dB (The distance from the front of device to the test instrument is 1m)
²
2、 When the icon
is mute), at this time the user should pay more attention to the patient.
3、 If occurrence of multiple alarm signal at the same time, the monitor will only show the high priority
alarm in the form of auditory alarm and light alarm. Meanwhile, all alarm signals information including
User Manual for Vital Signs Monitor
Tone Pitch
~400Hz
~500Hz
~500Hz
Table 5.2
displays on the screen and its color is red, that means the alarm volume is 0 (alarm
Beep Chain
10 beeps pause 3 sec.
3 beeps pause 5 sec.
Single beep
key to reset or pause the audible alarm indication