APPENDIX 1

State
Low voltage alarm
Spo2 alarm
Pulse rate alarm
Probe error alarm

APPENDIX 2

Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be
used in such environments:
IMMUNITY TEST
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below and should only
be used in such environments
IMMUNITY
IEC 60601
TEST
TEST LEVEL
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz
to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz
to 2.5 GHz
ENGLISH
Alarm condition delay
1s
330ms
330ms
16ms
IEC 60601
COMPLIANCE
TEST LEVEL
LEVEL
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
COMPLIANCE
LEVEL
3 V
3 V/m
Alarm signal generation delay
20ms
20ms
20ms
20ms
ELECTROMAGNETIC
ENVIRONMENT – GUIDANCE
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Power frequency magnetic fields should
be at levels characteristic of a typical lo-
cation in a typical commercial or hospital
environment.
ELECTROMAGNETIC
ENVIRONMENT – GUIDANCE
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance.
3.5
√P
d=
V
1
3.5
√P
d=
80 MHz to 800 MHz
E
1
7
√P
d=
800 MHz to 2.5 GHz
E
1
30