Hemostat Setup Screen Description:
² "Pressure": when you use hemostat function, you need to preset a cuff pressure for hemostasis. The
pressure is adjustable, and its adjusting limit is different for different patient category:
for neonates: preset range: 70~100 mmHg, default value: "90" mmHg;
for pediatrics: preset range: 80~130 mmHg, default value: "110" mmHg;
for adults: preset range: 80~180mmHg, default value: "140" mmHg.
If the pressure drops down slowly under 10mmHg compared with the preset value due to little air
leakage in the pneumatic system when time passes by, the monitor will re-inflate to maintain the cuff
pressure close to the preset pressure value.
Note: the unit of cuff pressure is the same with the NIBP unit in NIBP Setup.
² "Duration": After presetting the cuff pressure, you need to set the time period for maintaining the
preset pressure after inflation. "5, 6, 7,...120" minutes adjustable. The default value is "40" minutes.
If the set value is "xx" minutes, the monitor will count down from "xx" minutes automatically when
starting cuff inflation. When time is up, it will deflate automatically.
² "Alert Time": the alert time for reminding user that the operation of hemostat is going to be end after
this time period. 1 to 60 minutes adjusting range with 1 minute step, the default value is "5" minutes. If
the set value is "xx" minutes, the monitor will produce alarm sound until ending deflation when
counting down time reaches to "xx" minutes. The alarm type is high priority alarm. (For example: the
duration is 40 minutes, the alert time is 5 minutes, the alarm will ring for prompt when the duration
counting down to 5 minutes. The Prompt Info area starts to prompt: Hemo C-D 300 seconds. )
² "Start Hemostat": move the cursor to "Start Hemostat" and press "
Hemostat" button on the screen), then "Start Hemostat" becomes "Stop Hemostat" and meanwhile the
blood cuff starts being inflated; Pressing "Stop Hemostat" button can stop using this function. After
deflation, it will change to "Start Hemostat" again.
Safety instruction:
?
The NIBP calibration and Air leakage detection can only be carried on when the NIBP measurement is set to
mode "Manual".
NIBP setup screen description:
Short pressing NIBP setup "
User Manual for Vital Signs Monitor
Figure 4.14 Hemostat setup screen
" key to enter into NIBP setup screen, as shown in figure 4.15.
40
" key (or press "Start