When selecting a sensor, consider the patient's category, adequacy of perfusion, availability of probe site and
anticipated monitoring duration. Use only SpO
following table for SpO
probe.
SpO
SpO
SpO
3. If the neonate SpO
Y type sensor
 For proper placement on foot, place the sensors on the outside of the foot behind the pinky toe. Make sure
the sensor touch the skin closely, then secure the foot wrap with Velcro (see Figure 3.10B and 3.10C). Do not
over-tighten.
User Manual for Vital Signs Monitor
Figure 3.9 Finger clip SpO
probe information. Refer to Chapter 11.5 for the detailed instructions of each SpO
2
Probe
2
Finger Rubber Sensor(reusable)
2
Finger Clip Sensor(reusable)
2
sensor is used, please follow Figure 3.10 to connect.
2
Light source
Receiver
(round )
Figure 3.10A Neonate SpO
Pinky toe
Foot wrap
Ankle wrap
Left foot
(wrap goes under the foot)
sensor placement
2
probes provided by our company with this monitor. Read the
2
Patient Category
Adult
Adult
(square)
Foot wrap
Ankle wrap
sensor placement
2
Foot wrap
Ankle wrap
(wrap goes from top of the foot)
Figure 3.10B
20
Pinky toe
Right foot
2