When the device fails to try establishing wireless connection within 3 minutes,
the icon " " will become gray and the "Wireless" function is disabled
automatically. You have to enable it next time manually.
Notes: The pulse beep has the pitch-tone feature (when SpO
value
2
is higher than 90%), that means, the beeping tone changes according
to the SpO
value.
2

3.2 Alarm and alarm silence

When measuring, if SpO
value or pulse rate value exceeds the preset alarm
2
limit, the device will alarm automatically and the value which exceeds limit
on the screen will flash. The detailed information refers to chapter 4.
Apply for belowed methods to relief alarm sound when alarm event happens:
1. When SpO
value and PR value get normally.
2
2. Press Display Key to mute. If this alarm event continues, the oximeter will
resume alarm sound automatically 2 minutes later.
3. Remove the finger from the oximeter or SpO
probe.
2

3.3 Menu Screen

Figure 5 Menu Screen
Longtime press display key could enter the menu screen.
Menu screen description:
"Wireless": the wireless on-off button. Transmitting data to PC when it is on.
"on" and "off" can be optional. The factory default is "on".
"SpO
alm Lo": SpO
alarm: Lower limit.The user can modify the value
2
2
of 85~99, the step is "1", the default is 90.
"PR alm Hi": Pulse Rate alarm upper limit. The user can modify the value
of 100~240, the step is "5", the default is 120.
"PR alm Lo": Pulse Rate alarm lower limit. The user can modify the value
of 30~60, the step is "1", the default is 50.