estech COBRA 604-4810B Instrucciones De Uso página 44

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 357
430-11902-02 Rev N
Table 14-7 IEC EMC Specifications (Immunity)
Recommended separation distances between
portable and mobile RF communications equipment and the Estech
COBRA™ Electrosurgical HF Generator
1. The Estech COBRA™ Electrosurgical HF Generator is intended for use in an
electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Estech COBRA™ Electrosurgical
HF Generator can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Estech COBRA™ Electrosurgical HF
Generator as recommended below, according to the maximum output power
of the communications equipment.
Rated
maximum
output
power of
transmitter
W
0.01
0.1
1
10
50
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be determined using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
Separation distance according to frequency of transmitter M
150 kHz to 80 MHz
80 MHz to 800 MHz
3.5
d = [
d = [
] P
V
1
0.12
0.12
0.37
0.37
1.17
1.17
3.69
3.69
8.25
8.25
Page 44 of 411
800 MHz to 2.5 GHz
3.5
7
d = [
] P
E
E
1
0.23
0.74
2.33
7.38
16.50
] P
1
Tabla de contenido