ELECTRO-MAGNETIC COMPATIBILITY (EMC) SPECIFICATIONS
Table 14-5 IEC EMC Specifications (Emissions)
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The Estech COBRA™ Electrosurgical HF Generator is intended for use in the
electromagnetic environment specified below. The customer or the user of the
Estech COBRA™ Electrosurgical HF Generator should assure that it is used in
such an environment.
Emissions
Test
RF
emissions
CISPR 11
RF
emissions
CISPR 11
Harmonic
emissions
IEC 61000-
3-2
Voltage
fluctuations/
flicker
emissions
IEC 61000-
3-3
Electromagnetic Environment – Guidance
Compliance
The Estech COBRA™ Electrosurgical HF
Group 1
Generator must emit electromagnetic energy in
order to perform its intended function. Nearby
electronic equipment may be affected.
The Estech COBRA™ Electrosurgical HF
Class A
Generator is suitable for use in all establishments
other than domestic and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Class A
Complies
Page 40 of 411