430-11902-02 Rev N
chapter).
The ESU contains no user-serviceable parts; disassembly and attempted
repair by unqualified personnel may create a hazardous condition and
voids the warranty.
Cleaning/Disinfecting the ESU System
NOTE: Do not spray or pour liquids directly on the unit.
NOTE: The unit and/or accessories cannot be sterilized.
WARNING: Ensure Isopropyl Alcohol (IPA) is completely dry before
operating the unit.
CAUTION: Avoid caustic or abrasive cleaners
Guidelines
The following guidelines are recommended for cleaning the unit. It is the user's
responsibility to qualify any deviations from these processing methods.
1. Disconnect the unit or cart from the outlet before cleaning.
2. If the unit and/or accessories are contaminated with blood or other body fluids,
they shall be cleaned before the contamination can dry (within two hours of
contamination).
3. The outer surfaces of the unit and/or accessories shall be cleaned with 70% -
90% Isopropyl alcohol (IPA) wipes for a minimum of two minutes. Do not allow
fluids to enter the chassis.
4. Pay attention to all areas where fluids or soil may gather, such as under/ around
the handles or any tight crevices/ grooves.
5. Dry the unit and/or accessories with a dry, white lint-free cloth.
6. Conduct a final confirmation of the cleaning process by visually inspecting the
white cloth for remaining soil.
7. If soil remains on the white cloth, repeat steps 3 through 6.
8. After cleaning is complete, turn the unit on to perform Power On Self-Test
(POST). If any errors are received, contact AtriCure to begin return process.
Replacing the Main Fuses
CAUTION
1. Before replacing a fuse in the ESU, disconnect the mains power cord
WARNING
Do not open unit. Electrical shock hazard.
When replacing the fuse, ensure the fuse holder is re-
inserted in the correct orientation for AC Mains voltage.
30