temperature is greater than 120° C. Otherwise, it shows the measured
temperature.
Figure 5-4: TEMPERATURE Display and Buttons
TEMPERATURE SET Buttons
The TEMPERATURE SET (increase/decrease) Buttons
temperature that the ESU attempts to maintain during RF power delivery.
During Deliver Mode, the ESU automatically adjusts power, within the user-
selected upper power limit, to achieve the desired tissue setpoint
temperature.
SET/MEASURED LED Indicators
The SET and MEASURED LEDs (see Figure 5-5) are located to the right of
the numeric displays and indicate what is currently being displayed in the
POWER, TIMER and TEMPERATURE displays.
During RF DELIVER mode, the yellow MEASURED Indicator is normally lit
and the numeric displays indicate measured power, remaining time and
current temperature.
Whenever any of the up/down buttons for the POWER, TIMER or
TEMPERATURE is pressed, the yellow MEASURED LED turns off. The
green SET LED then turns on and the numeric display their respective
setpoint values.
Figure 5-5: SET and MEASURED LEDs