12.Tighten the sensor Clamping Screw to hold the sensor in place.
13.Observing the UP (upstream) and DN (downstream) orientation, attach the UP and DN
FUH1010 IP65 NEMA 7 Interface Detector Quick Start
Operating Instructions, 06/2014, A5E02428148-AA
procedure for the other sensor
①
Front View
②
Spring Clip (Not present on some models)
③
Sensor Clamping Screw
④
Spacer Bar
⑤
1012 Series Mounting Frame
Figure 3-9
Sensor Installation
cables to the sensors and make snug. Attach the other ends to the UP and DN terminals
of the flow meter (see figure below).
.
⑥
⑦
⑧
⑨
⑩
⑪
Commissioning
3.5 Sensor Installation
Repeat installation
Mounting Strap
Note: Optional 2nd Mounting Strap
shown. . Larger pipes over 76cm (30
inches) may need an additional
support.
Spacer Bar Platform and Clamping
Screw
Spacer Bar Reference Hole
Orientation for Single Beam Sensor at
9 o'clock position
Orientation for Dual Beam Sensor at
10 & 2 o'clock positions