Description
─
Electrosurgical unit code
─
Self-check
─
Last working condition storing
─
Electrical Class(EN60601-1)
─
MDD 93/42/EC Class
─
EN55011 (CISPR 11) Class (Class/Group)
─
Patient circuit
─
Duty Cycle (action / pause) in seconds
─
Output power control
─
Defibrillation-proof
EN
─
Equipotential binding
─
ABS cabinet
■= PRESENT
MAINTENANCE
General
No user adjustable parts are within the equipment, either for calibration or service pur-
poses.
The equipment housing must not be opened: the warranty is invalidated by any unau-
thorized entry into the unit. In the event any repair or adjustment work being necessary,
the whole equipment should be returned to the LED SpA Service Centre 04011 APRILIA
(LT) - ITALY, or to an other Authorized Centre, together with a description of the fault.
Maintenance work by the user is mainly the cleaning of the exterior of the cabinet, clean-
ing and sterilization of the accessory items and checking of the equipment before each
use. Carrying out function and safety check for verification of the parameters is demand-
ed to specialized technical people.
Cleaning of the Cabinet
Switch the equipment off completely and disconnect the mains supply before any clean-
ing is undertaken. Clean the outside of the cabinet with a damp cloth. No chemical
should be used; a mild non abrasive cleanser may be used when necessary.
Cleaning and Sterilization of the Accessories Items
The best thing to do is to use only one time use accessories and discard them after
use. Since some of the accessory items are to be used more than once it is mandatory
to clean carefully and sterilize those accessories before the new use. The best way to
clean and sterilize the reusable items is to follow the direction of the supplier of each
item. When original reusable accessories supplied by LED SpA are applied, the cleaning
by using deep cleanser and sterilization through steam sterilization at 121 °C / 134 °C
is recommended.
Guide to the Solution of the Problems
In case of problems before all it is advised to check for the correct installation of the unit
and for the correct connection of the accessories.
Problems
The equipment doesn't switch
on.
EN 18
Probable Cause
Interruption or absence of the
main feeding.
Verify the
connection of the main cable
SURGERY
PLUS
CDG10100.051
■
■
I CF
II b
2 / B
-F-
10 / 30
foot-switch
■
■
■
Solutione
Verificare il collegamento del
Verify the fuses and replace
them, where necessary, with
new ones of the proprie type.
MA505a_IT-EN-ES-FR-DE