Power Supply Module and Voltage Selector
Power supply module is the connection point of mains voltage feeding to the unit. This
module is provided with line fuses and the voltage selector.
WARNING: before switch on the unit, operator has to verify that requested mains volt-
age corresponds to the voltage available from the electrical net. (see chapter INSTAL-
LATION).
Power On-Off Switch
The POWER ON/OFF mechanical switch is used to control power to the equipment. To
power the equipment, press the switch in the direction of the 1. When the equipment
is powered the front panel will illuminate. Pressing the switch in the 0 direction will
cut power to the equipment, this operation allows it to be used as an emergency stop
switch, in the event of any fault.
TECHNICAL CHARACTERISTICS
Toll.
Description
─
Electrosurgical unit code
± 0%
Minimum presectable power
─
Level step
─
Digital level display
±20%
Maximum output power CUT (W)
±20%
Maximum output power CUT 1 (W)
±20%
Maximum output power CUT 2 (W)
±20%
Maximum output power COAG (W)
± 5%
Modulation factor CUT
± 5%
Modulation factor CUT 1 (@10 kHz)
± 5%
Modulation factor CUT 2 (@10 kHz)
± 5%
Modulation factor COAG (@10 kHz)
Crest Factor CUT
- 0.1 +0.2
± 0.3
Crest Factor CUT 1
± 0.3
Crest Factor CUT 2
± 0.3
Crest Factor COAG
± 10%
Working frequency (kHz)
± 15%
Maximum output voltage CUT (Vpp)
± 15%
Maximum output voltage CUT 1 (Vpp)
± 15%
Maximum output voltage CUT 2 (Vpp)
± 15%
Maximum output voltage COAG (Vpp)
± 0.5
Weight (Kg)
± 10
Size WxHxD (mm)
± 5%
Selectable power (Vac)
± 1%
Mains frequency (Hz)
─
Fuses (230Vac) 5x20 type TIMED
─
Fuses (115Vac) 5x20 type TIMED
± 10%
Electrical input power (VA)
± 10%
Electrical input current (A) 230Vac
± 10%
Electrical input current (A) 115Vac
MA505a_IT-EN-ES-FR-DE
SURGERY
PLUS
CDG10100.051
0
1
■
50 → 400Ω
45 → 400Ω
40 → 400Ω
40 → 400Ω
Pure 100%
Mod. 90%
Mod. 80%
Mod. 60%
1.5
1.8
2.1
2.3
600
1000
1000
1000
1000
2.5
190x85x239
115 –230
50-60
2x T1A
2x T2A
200
0.9
1.8
EN 17
EN