OPERATION MODE
Switch On
When switched on the electrosurgical unit performs automatically a test to establish the
correct operation of itself and of the connected accessories as well. In case anomaly is
found it is shown coded according to the chart codes brought in the chapter MAINTE-
NANCE. At the end of the control the equipment restores last use operational conditions.
Preselecting of the Deliverable Current
The deliverable current for the surgical operations can have preselected through push
button for:
Cut Current (CUT)
The best current for the cut is the pure sinusoidal wave without modu-
lation that means with duty-cycle 100%. Such current, proper for cut
without coagulation.
Coagulated-Cut Current (CUT 1)
The coagulated-cut current (CUT1) it is suited for coagulated cut when
a medium coagulation together the cut is desired.
Coagulated-Cut Current (CUT 2)
The coagulated-cut current (CUT2) it is suited for coagulated cut when
a more coagulation together the cut is desired.
Coagulation Current (COAG)
The modulated current COAG it is characterized by good property of
surface coagulation behaving at the time it probable production of es-
char and partial carbonization of the tissue. The advantage of this type
of coagulation resides in the rapidity with which the effect is gotten.