Checking of the Equipment Before Each Use
Each time the use of the electrosurgical equipment is planned a check of the most impor-
tant safety aspects has to be implemented considering at least the following:
●
Check the integrity of cords, connections, wires breakage, etc.
●
Assure that all the electrical equipment is properly grounded.
●
Assure that all the accessories that should be used are available and sterilized.
●
Check, by activating the power switch, the functioning of the emission sounds
warnings.
Function and Safety Check and Test
At least once a year, the biomedical engineering department or other qualified personnel
should do the following check and test:
●
Check of the connectors and mains supply cord conditions.
●
Visual check of the mechanical protections.
●
Check of the protections against the danger due to liquid's pouring, dripping,
EN
moisture, liquid's penetration, cleanliness, sterilization and disinfection.
●
Check of the Equipment's Data on the Label.
●
Check of the availability of the Instruction's Manual.
●
Check the functioning of the H.F. output controls.
●
Check the uniformity of the resistance through the surface of the patient plate.
●
Test the earth conductivity resistance.
●
Test the earth leakage current.
●
Test H.F. leakage current.
●
Control of the neuromuscular stimulation.
●
Control of the accuracy of the output power.
EN 20