Application
Transducer Placement
Locations
(See diagram below for reference)
A: Primary location
The preferred location and to be used when possible. The center of the tank generally
gives the most reliable readings as there are fewer obstructions, minimizing false echoes.
B: Alternate location
The alternate location and used if the centre of the tank is already in use or if the tank
roof cannot hold the transducer safely.
C: Poor location
The avoided location. The echoes will reflect away from the transducer face.
maintain full fluid
level for full or
offset calibration.
Do not allow
.
material to enter
the blanking zone.
empty level for B
location. Below this
level the echo reflects
away from the
transducer.
may require target to
obtain empty reading.
7ML19985HX62
XCT-8 Sanitary Transducer– INSTRUCTION MANUAL
A
B
run
discharge
C
span: distance between
beam angle
Empty and Full levels in
measurement process.
tank manufacturer's
empty level.
Page EN-7