wiring diagrams. Ensure that the switch is set properly before wiring to the inputs. The location of
the switch and the active/passive positioning are shown in Figure 2-70.
2.6.

PCB Jumper Positions

There are no jumpers on the Main board or any of the option boards in the IND246 terminal.
2.7.

SD Card Installation

The SD memory card can be used for additional storage in the Checkweighing and Counting
applications, and must be installed if the Vehicle Application is used. Figure 2-71 shows the
installation of an SD card into the socket on the edge of the IND246 main board.
2.8.

Capacity Label Instructions

The regulations in some locations require that the scale capacity and increment be shown on the
front of the terminal, near the display. To satisfy this requirement, a blue capacity label is included
with the terminal that must be completed and adhered to the front overlay.
The capacity label (shown in Figure 2-72) provides space for the Max, min, and e information for
both ranges for which the scale is programmed. If only one range is used, the unused portion of the
label may be cut off with scissors. Written information must be legible and a minimum of 2mm or
0.08 in. in height. A permanent marker should be used for this information.
64084463
|
03
|
12/2017
Figure 2-73: Location and Switch Settings for Discrete I/O Switch