15. Fold the individual strands of the inner braided shield back over the grommet. The individual
16. Separate the elements of the cable – the green, red and black wires, the inner drain wire, the
17. Cut the foil back to about an inch from the grommet. Snip lengthwise to allow it to fold back all
18. Insert the wires through the body of the gland and into the enclosure.
19. Press the plastic grommet into the body of the gland, then screw the outer part over it,
2-24
METTLER TOLEDO IND246 and IND246 POWERCELL Installation Manual
strands should be uniformly distributed around the outer surface of the grommet.
Figure 2-42: Inner Braided Shield Folded Back over Grommet
blue and white wires and their foil wrapper. The nylon string, used only during manufacture,
can be cut off.
around the grommet, and cut both foil and inner braided shield wires so that they cover the end
of the grommet, without covering the O-ring.
Figure 2-44: Inner Braided Shield Wires and Foil Trimmed to Length and Folded Back Over Grommet
tightening with an adjustable wrench. The end of the cable, with the wires and foil pressed into
Figure 2-43: Cable Components Separated
64084463
|
03
|
12/2017