Parameter
Target Pt
Rate
Target Va
IPAP
EPAP
Min PS
Max PS
Respiratory
Rate
Ti Max
Ti Min
Ti
Rise Time
*** 2-25 cm H
O for Europe and APAC
2
Modes
9
9
9 9 9 9
9 9 9 9 9 9
9 9 9 0–20 cm H
9 9 9 0–28 cm H
9 9 9
9 9 9
9
9 9 9
9
9 9
9 9 9 9 9
Range
Description
8–30 BPM
The rate input to the iVAPS
algorithm. This should be
Default: 15 BPM
set at the patient's actual
respiratory rate.
1–30 L/min
Used to determine the
amount of pressure support
Default: 5.2 L/min
required by the iVAPS
algorithm.
4–30 cm H
O
Sets the pressure which will
2
be delivered to the patient
Default:
when the device is triggered
10 cm H
O
2
into inspiration.
3–25 cm H
O***
Sets the pressure which will
2
be delivered to the patient
Default:
when the device is cycled
4 cm H
O
2
into expiration.
Dependent on IPAP.
O
Minimum pressure support
2
in iVAPS mode.
Default:
4 cm H
O
Dependent on EPAP.
2
O
Maximum pressure support
2
in iVAPS mode.
Default:
20 cm H
O
Dependent on EPAP and
2
Min PS.
5–50 BPM
Sets the breaths per minute
(BPM) or 'backup' rate.
Default: 10 BPM
0.3–4.0 sec
Sets the maximum limit on
the time the device spends
Default: 2 sec
in IPAP .
Dependent on Respiratory
Rate.
0.1–4.0 sec
Sets the minimum limit on
the time the device spends
Default: 0.3 sec
in IPAP.
Dependent on Ti Max.
0.1–4.0 sec
Sets the duration of
inspiration in timed breath.
Default: 2 sec
Dependent on Respiratory
Rate.
Min /
Sets the time taken for the
100–900 ms
device to reach to IPAP .
Default: 150 ms
Dependent on Ti Max and Ti.
17
English