Purpose
en
12 / 412
WARNING!
Falls and crashes
Place the device on a stable and level surface. Cables must be laid safely. Tubes
between the device and the patient must not create any obstruction.
WARNING!
Function test
The function test must be performed prior to each device use.
WARNING!
Obvious defects
Never use the system if it has suspected or confirmed defects, especially if these in-
volve the power plugs or the mains power supply connection cables. In this case
have the device repaired by authorized service personnel.
WARNING!
System error
Do not use the Aquilex® System if a defect is suspected or detected during the
function test. This also applies to any obvious defects, especially defects on the
power connector or plug and power cord.
WARNING!
Device setup
Device should be positioned outside of the sterile area in such a way that
• it can be easily disconnected,
• it is easy to use and switch off and on,
• it allows an easy monitoring of the display values, device functions, and access
to the control elements.
WARNING!
ON/OFF switch
For electrical safety reasons, do not touch the patient and the ON/OFF switch at
the same time.
WARNING!
Disconnect the power cord
Pressing the ON/OFF switch does not disconnect the system from the wall power
outlet. This requires pulling the power cord located in the rear of the system.
WARNING!
Unplug the power cord from the system before checking the fuse.
WARNING!
Technique and procedures
Only the physician can evaluate the clinical factors involved with each patient and
determine if the use of this device is indicated. The physician must determine the
specific technique and procedure that will accomplish the desired clinical effect.