Section 4 Operation
Changing the
Setpoint
Setpoint Value Window
Setpoint Value Window
Setpoint Value Window
4-40
Note You can not adjust the setpoint closer than 0.1°C to either of
the fluid's system limits, see Fluids Type in this Section, or beyond the
circulator's temperature range.
The setpoint can be changed with the circulator running or not.
The Setpoint is the desired process fluid temperature. The circulator
can store up to five setpoints,
changing stored setpoint values is discussed later in this Section.
Use the navigation arrows and move to the setpoint window and then
press
to highlight it and the setpoint value as shown below.
Setpoint Window
Use the up and down navigation arrows to bring up the desired
SP1
setpoint,
through
The Setpoint Value Window now indicates the corresponding setpoint's
stored value.
If desired, you can change the displayed setpoint value by using the
navigation arrows to highlight the Setpoint Value Window and then
. The right-most digit will have a cursor beneath it.
pressing
Use the left and right arrows to move the cursor to the desired digit
and then use the up and down arrows to change the value. Once all the
desired changes are made, press
Note Using this procedure also changes the setpoint's stored value.
SP1
through
SP1
20.0°C
SP5
, and then press
SP2
35.0°C
35.0°C
to save the change.
SP5
. The procedure for
Menu
In
24.2°C
.
Menu
In
24.2°C
Thermo Scientific