Section 3 Installation

Filling Requirements

3-34
Ensure the reservoir drain port on the front of the bath is closed and
that all plumbing connections are secure. Also ensure any residue is
thoroughly removed before filling the bath.
Before using any fluid refer to the manufacturer's SDS and EC
safety data sheets for handling precautions.
To avoid spilling, place your containers into the bath before filling.
With a low level WARNING the circulator continues to run, with a
FAULT the circulator shuts down the refrigeration, pump and heater,
see Section 6. The low level warning is at approximately 43 mm (1 3/4")
below the top, the low level fault is at approximately 57 mm (2 1/4").
Oil-based fluids expand when heated. Fill the reservoir only to the
minimum level. Monitor the fluid level whenever heating the fluid.
When pumping to an external system, keep extra fluid on hand to
maintain the proper level in the circulating lines and the external system.

Draining

Before draining any fluid refer to the manufacturer's SDS and EC
safety data sheets for handling precautions.
Ensure the fluid is at a safe handling temperature, ~40°C. Wear
protective clothing and gloves.
• place a suitable vessel underneath the drain. If desired, attach an
• slowly turn the drain plug until flow is observed.
Turning the drain cap more than 1
cap and fitting coming off the bath.
In this case screw the drain fitting back into the bath. Attaching the cap
onto the fitting aids in installation. If required, contact us for additional
information.
8 mm id tube on the drain.
Installed Drain Fitting with Cap Removed
/
turns will result in the drain
1
2