Section 4 Operation
5. To pause the program at the current tempera-
ture, highlight the pause program symbol
press
.
To restart the program press
Ramp
25.00°C
Interfaces
is used to enable/configure the optional serial communications feature.
Interfaces
1. With
highlighted, press
display the list of parameters.
Highlight the desired parameter and press
view/change the available options.
Serial Type
Baud
Parity
Data Bits
Menu
4-50
and
again.
Menu
In
24.89 °C
to
6. When the program is complete the circulator
Shuts Down
either
setpoint.
Ramp
25.00°C
When
is pressed after the ramp program is
complete the refrigeration, pump and circulator
shut off.
Available options:
Serial Type
Off, RS232, RS485 or Analog IO
Baud
38400, 19200, 9600, 4800, 2400,
to
1200, 600 or 300
Parity
None, Odd or Even
Data Bits
8 only
Stop Bits
1 or 2
Address
(Displayed for RS485 only)
Supported protocols: AC, Standard, NC, Namur
See the Appendix for additional information.
Note Keypad operation is still available with serial
communications enabled.
Maintains
or
the last
Menu
In
25.00°C
Thermo Scientific